FDA Adverse Event Malfunction Summary report: N

GREENLIGHT HPS BPH FIBER OPTIC

MDR report key: 2974370 · Received February 19, 2013

Report

Report Number
2937094-2013-00250
Event Type
Malfunction
Date Received
February 19, 2013
Date of Event
November 15, 2012
Report Date
December 26, 2012
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE COMPONENT CODES FIBER AND CAP REFER TO THE RESULTS CODE THERMAL PROBLEM. FIBER ANALYSIS: THE FIBER CAP WAS FOUND TO BE DRILLED THROUGH. THE CAP WAS ALSO FOUND TO EXHIBIT DETRITUS, CHAR AND DEVITRIFICATION. THE FIBER CAP CONDITION WOULD PREVENT PROPER FIBER FUNCTION; LIKELY RESULTING IN A DIFFUSE BEAM AND REDUCED TISSUE VAPORIZATION EFFICIENCY. THE DRILLED THROUGH FIBER CAP CONDITION MAY ALSO RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT 166,058 JOULES, WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, DECREASED TISSUE VAPORIZATION EFFICIENCY WAS OBSERVED. THE CASE WAS COMPLETED USING A SECOND FIBER WITHOUT FURTHER ISSUE. THERE WAS NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72434 GREENLIGHT HPS BPH FIBER OPTIC POWERED SURGICAL LASER INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS 10-2090 211H

Patients

Seq Age Sex Outcome Treatment
1 GREENLIGHT HPS LASER SYSTEM| ACCESSORIES