SYSTEM 6 ASEPTIC HOUSING ASSY
Report
- Report Number
- 0001811755-2013-00360
- Event Type
- Malfunction
- Date Received
- February 22, 2013
- Date of Event
- February 7, 2013
- Report Date
- February 11, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE BATTERY HOUSING WAS CONFIRMED TO BE MISSING THE DELRIN BALL CAUSING THE LATCH OF THE BATTERY HOUSING TO NOT PROPERLY LOCK IN PLACE. BASED ON SIMILAR COMPLAINTS, POSSIBLE CAUSES INCLUDE NORMAL WEAR AND TEAR OR MECHANICAL DAMAGE THAT CAUSED THE BALL TO SHATTER.
THE DEVICE IS AVAILABLE FOR EVALUATION BUT HAS NOT YET BEEN RECEIVED. ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE DEVICE IS RECEIVED AND THE QUALITY INVESTIGATION IS COMPLETE. THE DEVICE IS AVAILABLE FOR EVALUATION BUT HAS NOT YET BEEN RECEIVED. ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE DEVICE IS RECEIVED AND THE QUALITY INVESTIGATION IS COMPLETE.
IT WAS REPORTED THAT THE SYSTEM 6 ASEPTIC BATTERY HOUSING OPENED DURING A PROCEDURE. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AS A RESULT OF THIS EVENT AND NO ADVERSE CONSEQUENCES.
IT WAS REPORTED THAT THE SYSTEM 6 ASEPTIC BATTERY HOUSING OPENED DURING A PROCEDURE. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AS A RESULT OF THIS EVENT AND NO ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 78732 | SYSTEM 6 ASEPTIC HOUSING ASSY | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | KIJ | STRYKER INSTRUMENTS-KALAMAZOO | 11244 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |