FDA Adverse Event Other Summary report: N

STOCKERT HEATING-COOLER SYSTEM 3T

MDR report key: 2974243 · Received February 14, 2013

Report

Report Number
1718850-2013-00018
Event Type
Other
Date Received
February 14, 2013
Date of Event
January 17, 2013
Report Date
January 18, 2013
Manufacturer
SORIN GROUP DEUTSCHLAND
Product Code
DWC
PMA / PMN Number
K052601
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SORIN GROUP (B)(4) MANUFACTURERS THE STOCKERT HEATER-COOLER SYS 3T. THE INCIDENT OCCURRED IN (B)(6). SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE PT SIDE OF THE 3T HEATER/COOLER WOULD NOT WARM ABOVE 11 DEGREES DURING A PROCEDURE. A BACK-UP HEATER/COOLER WAS USED TO COMPLETE THE CASE WITH NO FURTHER ISSUES. THERE WAS NO REPORT OF PT INJURY. A SORIN GROUP FIELD SVC REP WAS DISPATCHED TO INVESTIGATE. WHILE AT THE FACILITY THE FIELD SVC REP WAS ABLE TO REPRODUCE THE ISSUE AND FOUND THAT THE VOLTAGE TO THE HEATER ELEMENT WAS OUT OF SPECIFICATION. THE AC MAINS BOARD WAS REPLACED AND SUBSEQUENT TESTING CONFIRMED THAT THE ISSUE WAS RESOLVED. THE INVESTIGATION IS ONGOING. A F/U REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

SORIN GROUP (B)(6) RECEIVED A REPORT THAT THE PT SIDE OF THE 3T HEATER/COOLER WOULD NOT WARM ABOVE 11 DEGREES DURING A PROCEDURE. A BACK-UP HEATER/COOLER WAS USED TO COMPLETE THE CASE WITH NO FURTHER ISSUES. THERE WAS NO REPORT OF PT INJURY.