STOCKERT HEATING-COOLER SYSTEM 3T
Report
- Report Number
- 1718850-2013-00018
- Event Type
- Other
- Date Received
- February 14, 2013
- Date of Event
- January 17, 2013
- Report Date
- January 18, 2013
- Manufacturer
- SORIN GROUP DEUTSCHLAND
- Product Code
- DWC
- PMA / PMN Number
- K052601
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
SORIN GROUP (B)(4) MANUFACTURERS THE STOCKERT HEATER-COOLER SYS 3T. THE INCIDENT OCCURRED IN (B)(6). SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE PT SIDE OF THE 3T HEATER/COOLER WOULD NOT WARM ABOVE 11 DEGREES DURING A PROCEDURE. A BACK-UP HEATER/COOLER WAS USED TO COMPLETE THE CASE WITH NO FURTHER ISSUES. THERE WAS NO REPORT OF PT INJURY. A SORIN GROUP FIELD SVC REP WAS DISPATCHED TO INVESTIGATE. WHILE AT THE FACILITY THE FIELD SVC REP WAS ABLE TO REPRODUCE THE ISSUE AND FOUND THAT THE VOLTAGE TO THE HEATER ELEMENT WAS OUT OF SPECIFICATION. THE AC MAINS BOARD WAS REPLACED AND SUBSEQUENT TESTING CONFIRMED THAT THE ISSUE WAS RESOLVED. THE INVESTIGATION IS ONGOING. A F/U REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETE.
SORIN GROUP (B)(6) RECEIVED A REPORT THAT THE PT SIDE OF THE 3T HEATER/COOLER WOULD NOT WARM ABOVE 11 DEGREES DURING A PROCEDURE. A BACK-UP HEATER/COOLER WAS USED TO COMPLETE THE CASE WITH NO FURTHER ISSUES. THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66165 | STOCKERT HEATING-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | SORIN GROUP DEUTSCHLAND | 16-02-85 | NA |