FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 2974146 · Received February 19, 2013

Report

Report Number
9615050-2013-00242
Event Type
Malfunction
Date Received
February 19, 2013
Date of Event
January 1, 2013
Report Date
January 24, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K041550
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND AT POWER UP, THE DEVICE DISPLAYED A WHITESCREEN ERROR AND THE DEVICE SOUNDED AN AUDIBLE ALARM FROM THE SECONDARY BEEPER. THIS WAS DUE TO A SOFTWARE ERROR. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY THE HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED A WHITESCREEN ERROR. THE PUMP WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, NO SPECIFIC DETAILS WERE AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY AFTER THE DEVICE WAS POWERED ON, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72539 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA