AXON ECLIPSE CONTROLLER
Report
- Report Number
- 1045254-2013-00120
- Event Type
- Malfunction
- Date Received
- February 19, 2013
- Report Date
- November 27, 2012
- Manufacturer
- XOMED MFG JACKSONVILLE
- Product Code
- GWF
- PMA / PMN Number
- K061639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE STIM BOARD FAILED THE STIMULATION TESTS AND WAS REMOVED/REPLACE. IN ADDITION, UNRELATED TO THE STIM BOARD FAILURE, THE CONNECTOR BOARD PANEL WAS REPLACED DUE TO BROKEN EMG SPEAKER PINS. THE MANUFACTURING DATE IS UNOBTAINABLE. RECORDS PRIOR TO (B)(4) 2011, REMAIN WITH THE ORIGINAL DEVICE MANUFACTURER. THE NIM-ECLIPSE SYSTEM NEUROVASCULAR WORKSTATION IS INTENDED FOR USE TO MONITOR SENSORY AND MOTOR PATHWAYS. IF THE SYSTEM FAILS TO PERFORM AS INTENDED, (ESPECIALLY TO EFFECT OR DEFECT ELECTROMYOGRAPHIC (EMG) ACTIVITY) IT PRESENTS THE POTENTIAL FOR TEMPORARY OR PERMANENT DAMAGE TO THE NERVE THAT IS PRESENT. THE ECLC ECLIPSE SYSTEM CONTROLLER PROVIDES HIGH SPEED DIGITAL DATA PROCESSING, STIMULATION GENERATION AND AUDIO PROCESSING OF EMG ACTIVITY AND ALSO SUPPLIES SWITCHED AC POWER TO THE COMPUTER. THE CONTROLLER CONNECTS TO THE COMPUTER VIA THE HIGH SPEED (B)(4) INTERFACE. WHEN INFORMATION SUGGESTS THAT THE NIM ECLIPSE EQUIPMENT HAD THE POTENTIAL TO NOT STIMULATE TO AFFECT ELECTROMYOGRAPHY (EMG), OR TO DETECT EMG, IT IS ASSUMED THAT THE FAILURE WAS DEVICE-RELATED AND MAY HAVE GONE UNDETECTED. IN THIS CASE, THERE IS NO INFORMATION TO SUGGEST AN EVENT COULD NOT BE INTERPRETED FALSELY - THE REPORT IS INCONCLUSIVE AS TO THE CAUSE OF FIELD OBSERVATION. THEREFORE, WITHOUT INFORMATION TO REASONABLY SUGGEST SERIOUS INJURY, OR THAT MEDICAL INTERVENTION WAS REQUIRED TO PRECLUDE SERIOUS INJURY, WE ARE FILING THIS REPORT AS A PRODUCT PROBLEM. THIS PRODUCT WAS BEING USED FOR TREATMENT, NOT DIAGNOSIS.
THE MANUFACTURER HAS REFINED THE CRITERIA FOR MAKING MDR DECISIONS. UPON A RETROSPECTIVE REVIEW, THE FOLLOWING EVENT IS NOW BELIEVED TO BE REPORTABLE: A CUSTOMER RETURNED AN ECLIPSE CONTROLLER - THE COMPLAINT IS UNKNOWN, UNIT RECEIVED WITHOUT ANY INFORMATION. THERE WAS NO SUGGESTION OF PATIENT INJURY OR INVOLVEMENT. TESTING/REPAIR FOUND A FAILED STIMULATION BOARD, WHICH HAS THE POTENTIAL TO CAUSE OR CONTRIBUTE TO PATIENT INJURY (LOSS OF STIMULATION DURING USE LEADING TO A FALSE NEGATIVE) - AS A SUDDEN FAILURE IS NOT CONSISTENTLY ACCOMPANIED BY WARNING/ERROR MESSAGES DURING A PROCEDURE. AT THIS TIME WE ARE UNABLE TO CONFIRM WHETHER THE CUSTOMER RECEIVED ANY ERROR MESSAGES OR ALERTS OF THE FAULT. THIS REPORTED EVENT HAS NO REFERENCE TO AN ALARM OR WARNING, THEREFORE, IT MUST BE ASSUMED THAT AN ALARM OR WARNING WENT UNDETECTED OR DID NOT OCCUR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 72656 | AXON ECLIPSE CONTROLLER | GWF - STIMULATOR, ELECTRICAL, EVOKED RESPONSE | GWF | XOMED MFG JACKSONVILLE | ECLC | NOT PROVIDED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |