FDA Adverse Event Malfunction Summary report: N

PORTABLE VIDEO MONITOR, GVL-2000

MDR report key: 2974098 · Received February 19, 2013

Report

Report Number
9615393-2013-00101
Event Type
Malfunction
Date Received
February 19, 2013
Date of Event
January 21, 2013
Report Date
January 21, 2013
Manufacturer
VERATHON MEDICAL (CANADA) ULC
Product Code
CCW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE RECEIVED THE DEVICE FOR FURTHER EVALUATION. THROUGH VISUAL INSPECTION OF THE LCD SERVICE CONFIRMED THAT THE IMAGE IS VERY DARK. SERVICE REPLACED THE LCD AND IMAGE IS BACK TO NORMAL. THE LCD WAS THE ROOT CAUSE OF THE REPORTED EVENT.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED VERATHON INC. REGARDING A PORTABLE VIDEO MONITOR, GVL-2000 SCREEN THAT IS NOT LIGHTING UP ALL THE WAY AND THE IMAGE IS VERY DARK AND DIFFICULT TO VIEW. DURING TROUBLESHOOTING OF THE REPORTED EVENT THE CUSTOMER TRIED SEVERAL BLADES WITH THE SAME MONITOR AND THE IMAGE IS THE SAME. NO INJURY TO PATIENT WAS REPORTED WITH THIS EVENT. THE DEVICE WAS RETURNED FOR FURTHER EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72888 PORTABLE VIDEO MONITOR, GVL-2000 NONE CCW VERATHON MEDICAL (CANADA) ULC 0231-0003