FDA Adverse Event
Malfunction
Summary report: N
CORE CONSOLE WITH INTEGRAL IRRIGATION PUMP
MDR report key: 2974068
·
Received February 22, 2013
Report
- Report Number
- 0001811755-2013-00350
- Event Type
- Malfunction
- Date Received
- February 22, 2013
- Date of Event
- January 25, 2013
- Report Date
- January 28, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CORE CONSOLE WAS CLEANED, LUBRICATED, ADJUSTED AND RETURNED TO CIRCULATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT A STRYKER LOANER CORE CONSOLE WITH IRRIGATION PUMP FAILED A CHASSIS EARTH LEAKAGE TEST. THE EVENT OCCURRED DURING A ROUTINE TESTING PRIOR TO CIRCULATION. THERE WAS NO PATIENT INVOLVEMENT, NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 78915 | CORE CONSOLE WITH INTEGRAL IRRIGATION PUMP | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |