FDA Adverse Event Malfunction Summary report: N

I-STAT ACT KAOLIN CARTRIDGE

MDR report key: 2974022 · Received February 22, 2013

Report

Report Number
2245578-2013-00020
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
January 7, 2013
Report Date
January 7, 2013
Manufacturer
ABBOTT POINT OF CARE
Product Code
JBP
PMA / PMN Number
K023582
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

APOC INCIDENT #(B)(4). THE INVESTIGATION WAS COMPLETED ON (B)(4) 2013. RETAIN TESTING WAS COMPLETED AND PRODUCT IS PERFORMING AS INTENDED. HOWEVER, TESTING OF CUSTOMER RETURNED CARTRIDGES SHOW A HIGHER RATE OF QUALITY CHECK CODES (QCC) THAN EXPECTED. GATHERED PROCESS TEST DATA INCLUDING FINISHED GOODS, RETAIN AND RETURN DATA FOR ACT KAOLIN S12276 SHOWS THE INITIAL OSCILLATION MEASUREMENT CAN BE A PREDICTOR FOR INCREASED RATES OF CODES LIKE QCC CODE 31. THE ISSUE IS NOT UNIFORMLY DISTRIBUTED ACROSS LOT S12276.

Description of Event or Problem · 1

ON (B)(6) 2012, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING ACT KAOLIN CARTRIDGES THAT YIELDED MULTIPLE QUALITY CHECK CODES 31 AND QCC 22 WITH ACT-K CARTRIDGE LOT # S12276 ON PATIENTS. PER I-STAT SYSTEM MANUAL (ART: 714260-00N) CODE 31 UNABLE TO POSITION SAMPLE/ CARTRIDGE ERROR: THE ANALYZER DID NOT DETECT MOVEMENT OF SAMPLE ACROSS THE SENSORS. THIS COULD BE DUE TO A CLOT IN THE SAMPLE (ESPECIALLY IN NEONATES), TO NOT CLOSING THE SNAP CLOSURE ON THE CARTRIDGE, OR TO AN ABERRANT CARTRIDGE. CODE 22 CARTRIDGE ERROR: THESE CODES OCCUR ONLY FOR COAGULATION CARTRIDGES IF THE MIXING OF THE SAMPLE AND REAGENT IS COMPROMISED. THIS CAN BE CAUSED BY AN INSUFFICIENT OR CLOTTED SAMPLE, OR BY AIR BUBBLES IN THE SAMPLE. ABBOTT POINT OF CARE HAS DETERMINED THE EVENT AS STATED BY THE CUSTOMER, A POTENTIAL FOR A SIGNIFICANT DELAY BASED ON THE AVAILABLE INFORMATION. THE INFORMATION PRESENTS A HIGHER RATE OF QUALITY CHECK CODES THAN EXPECTED. THE CUSTOMER DOES NOT PROVIDE PATIENT INFORMATION. BASED ON THE INFORMATION AT THIS TIME THERE WERE NO INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78718 I-STAT ACT KAOLIN CARTRIDGE ACT KAOLIN CARTRIDGE JBP ABBOTT POINT OF CARE NA S12276

Patients

Seq Age Sex Outcome Treatment
1