FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 2973986 · Received February 13, 2013

Report

Report Number
1028232-2013-00250
Event Type
Injury
Date Received
February 13, 2013
Date of Event
November 30, 2012
Report Date
January 29, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
DTB
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RV LEAD HAD DISLODGED. A REVISION PROCEDURE WAS PERFORMED, IN WHICH THE SAME LEAD WAS REPOSITIONED. IT WAS INDICATED THAT THE PATIENT WAS DOING FINE AND THAT THE PATIENT WAS NOT PACER DEPENDENT. AT THIS TIME, NO ADVERSE PATIENT EFFECTS WERE REPORTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64328 DEXTRUS 4137 PACER LEAD DTB BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization