FDA Adverse Event Injury Summary report: N

ARTHROSCOPY BLADE

MDR report key: 2973908 · Received February 14, 2013

Report

Report Number
MW5029039
Event Type
Injury
Date Received
February 14, 2013
Date of Event
February 6, 2013
Report Date
February 14, 2013
Manufacturer
STRYKER ENDOSCOPY
Product Code
HRX
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

I SUBMITTED A MEDWATCH REPORT ON (B)(6) RELATED TO APPARENT METALLIC FRAGMENTS FROM AN ARTHROSCOPIC CUTTING BLADE. WE HAVE SUBSEQUENTLY CONFIRMED THROUGH SCANNING ELECTRON MICROSCOPY AND ENERGY-DISPERSIVE X-RAY SPECTROSCOPY (SEM/EDS) THAT THE FRAGMENTS IDENTIFIED IN THE SURGICAL PATHOLOGY SPECIMEN WERE STAINLESS STEEL AND THAT THE SPECTROSCOPY MATCHES THE COMPOSITION OF THE CUTTING BLADE. THE ELECTRON MICROSCOPY ALSO INDICATES FRAGMENTS ON THE TEETH OF AN UNUSED CUTTING BLADE.

Description of Event or Problem · 1

PATIENT HAD A PRE-OPERATIVE DIAGNOSIS OF RIGHT SHOULDER LABRAL TEAR. DURING AN ARTHROSCOPIC SHOULDER PROCEDURE RIGHT SHOULDER ARTHROSCOPY WITH LIMITED LABRAL DEBRIDEMENT, LIMITED SYNOVECTOMY SURGICAL LIGHTING AND PICTURES INDICATE SHINY REFLECTIVE PARTICLES NOT TRADITIONALLY OBSERVED DURING THIS PROCEDURE WHICH CAUSED CONCERN TO THE SURGEON TO WARRANT STOPPING THE PROCEDURE AND FURTHER ANALYSIS. THE FOLLOWING DETAILS INFO FROM THE SURGICAL REPORT: WHILE SHAVING THE LABRAL EDGE, SMALL AREAS OF DARK DISLOCATION WERE NOTED REMAINING ON THE TISSUE. WHEN VIEWED CLOSELY, THEY OCCASIONALLY REFLECTED LIGHT FROM THE SCOPE. THE FINDINGS WERE HIGHLY SUSPICIOUS THESE WERE MICROSCOPIC METALLIC FOREIGN BODIES ARISING FROM THE SHAVER BLADES. DEBRIDEMENT WAS DISCONTINUED AND THE SHAVER AND TIP QUARANTINED UNTIL BIOMED COULD COLLECT THEM. AS MUCH OF THE MATERIAL AS POSSIBLE WAS REMOVED WITH PUNCH FORCEPS. A COUPLE OF VERY SMALL FRAGMENTS WERE ADHERENT TO THE SURFACE OF THE GLENOID ARTICULAR CARTILAGE AND WERE NOT REMOVABLE WITH GENTLE ATTEMPTS TO ASPIRATE, OR GRASP THEM. IT WAS FELT THAT THE DEFECTS CREATED FOLLOWING AGGRESSIVE ¿DIGGING¿ TO REMOVE THE FRAGMENTS WOULD HAVE A GREATER LIKELIHOOD OF LONG-TERM IMPAIRMENT, SO THE FRAGMENTS WERE LEFT IN SITU.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66170 ARTHROSCOPY BLADE DISPOSABLE ARTHROSCOPY BLADE HRX STRYKER ENDOSCOPY 0375-544-000 12306CE2
66171 STRYKER FORMULA HANDPIECE HANDPIECE HWE STRYKER ENDOSCOPY

Patients

Seq Age Sex Outcome Treatment
1 25 YR Other