FDA Adverse Event Injury Summary report: N

ALARIS CC GUARDRAILS PLUS SYRINGE PUMP

MDR report key: 2973890 · Received February 15, 2013

Report

Report Number
MW5029040
Event Type
Injury
Date Received
February 15, 2013
Date of Event
January 29, 2012
Report Date
January 31, 2013
Manufacturer
CAREFUSION
Product Code
FRN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

A PATIENT IN SEPTIC SHOCK RECEIVING A NORADRENALINE (NOREPINEPHRINE) - INFUSION VIA AN ALARIS CC GUARDRAILS PLUS SYRINGE PUMP (CAREFUSION, SAN DIEGO CA, USA) RECEIVED AN INAPPROPRIATE DOSAGE OF NORADRENALINE RESULTING IN A SHARP RISE IN ARTERIAL BLOOD PRESSURE. ACCORDING TO CONVENTION AND INTERNATIONAL LITERATURE THE DOSAGE OF CATECHOLAMINES AND VASOPRESSORS ON OUR ICU ARE HANDLED IN UG/KG/MIN. UNTIL A RECENT SOFTWARE UPDATE THE DISPLAY OF THE SYRINGE PUMPS SHOWED ALL NUMBERS BEFORE AND AFTER THE DECIMAL POINT IN EQUAL FONT SIZE. NOW, THE NUMBERS AFTER THE DECIMAL POINT ARE DISPLAYED IN A DISTINCT SMALLER FONT SIZE WHICH MEANS THAT THEY ARE NOT ANYMORE LEGIBLE FROM A DISTANCE OF MORE THAN ONE METER. IN THIS CASE, AFTER A WRONG ADJUSTMENT OF THE SETTING OF THE PUMP, THE ATTENDING PHYSICIAN WAS NOT AWARE OF THE MIS-PROGRAMMING OF THE PUMP BECAUSE HE MISREAD THE VALUES ON THE SYRINGE-PUMP DISPLAY. THIS SERIOUS SAFETY PROBLEM HAS RECENTLY BEEN REPORTED TO MFR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69382 ALARIS CC GUARDRAILS PLUS SYRINGE PUMP SYRINGE PUMP FRN CAREFUSION

Patients

Seq Age Sex Outcome Treatment
1 76 YR Life Threatening