FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 2973857 · Received February 22, 2013

Report

Report Number
3005075853-2013-00800
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
February 4, 2013
Report Date
February 5, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION SENT; PENDING ANSWERS. LIVER, SPLEEN AND KIDNEY HAVE LITTLE CONSEQUENCE WHEN BURNED. HOLLOW MUSCULAR ORGANS LIKE STOMACH, BLADDER AND UTERUS ARE SLIGHTLY LESS TOLERANT THAN SOLID ORGANS. AND THIN WALLED HOLLOW STRUCTURES SUCH AS SMALL BOWEL, COLON AND ESOPHAGUS ARE AT MOST RISK. WITH INTERNAL BURNS, I AM REQUIRED TO OBTAIN FOLLOW-UP FROM THE SURGEON, AS THE PATIENT RECEIVED A BOWEL BURN AND THE OUTCOME OF THE BURN MAY NOT BE IMMEDIATELY KNOWN FOR 5-7 DAYS POST-OP, COULD YOU PLEASE ADVISE: HAS THE PATIENT HAD ANY SYMPTOMS? WHAT IS THE CURRENT PATIENT CONDITION? HAS THE SURGEON BEEN IN-SERVICED ON HEAT MANAGEMENT? IS THE SURGEON AWARE OF THE WARNINGS IN THE IFU AS TO HEAT?

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GEN04 AND IT DID ACTIVATE. DURING FUNCTIONAL TESTING, THE DEVICE WAS ACTIVATED FOR ABOUT 1 MINUTE WITH NO ABNORMAL HEAT OBSERVED. DURING AND FOLLOWING ACTIVATION IN TISSUE, THE INSTRUMENT BLADE, CLAMP ARM, AND DISTAL 7 CM OF THE SHAFT MAY BE HOT. AVOID UNINTENDED CONTACT WITH TISSUE, DRAPES, SURGICAL GOWNS, AT ALL TIMES. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOW ANTERIOR RESECTION, IT WAS FOUND THAT WHITE THERMAL DAMAGE (ABOUT 1CM IN LENGTH) ON THE SMALL BOWEL THOUGH THE CLAMP ARM TOUCHED THE SMALL BOWEL FOR A SHORT TIME WHEN THE DEVICE WAS USED FOR ABRASION OF THE FAT ON THE ABDOMINAL WALL. THERE WAS NO SMALL-BOWEL PERFORATION AND A SUTURE FOR THE THERMAL DAMAGE WAS PERFORMED TO COMPLETE THE CASE VIA ALREADY SET SMALL INCISION FOR AN ANASTOMOSIS WITHOUT CONVERTING THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THERE WAS NO THERMAL DAMAGE ON THE SURGEON. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78050 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK J90W7R

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE