FDA Adverse Event
Other
Summary report: N
LIGHTSHEER ( ALL MODELS)
MDR report key: 2973803
·
Received August 7, 2007
Report
- Report Number
- 2914019-2007-00024
- Event Type
- Other
- Date Received
- August 7, 2007
- Date of Event
- July 2, 2007
- Report Date
- August 7, 2007
- Manufacturer
- LUMENIS, INC.
- Product Code
- GEX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PER SERVICE TECH, REGULATORY INSPECTION NOTED DIRTY TOUCHSCREEN AND ENERGY METER GLASS. HANDPIECE TIP DIRTY AND LOOKS CLOUDED. PERFORMED INITIAL FINAL AND CONFIRM ENERGY OUTPUT. FIRED AT 37J WHICH WAS AT 40MS. THE DISPLAY GASKET WAS DISLOCATED AND DIRTY, REPLACED THE GASKET. THE ENERGY METER GLASS HAD THICK COATING OF SUBSTANCE. CLEANED TOUCHSCREEN, ENERGY METER GLASS, AND CLEAN TIP. PERFORMED ENERGY METER CALIBRATION. PERFORMED NECESSARY RELIABILITY IMPROVEMENTS AND FINAL TEST. ROOT CAUSE TOUCHSCREEN AND METER GLASS DIRTY.
Description of Event or Problem · 1
ONE PATIENT HAS BEEN BURNED, NAME IS (B)(6) AND I AM THE AESTHETIC NURSE. THE PROCEDURE WAS LASER HAIR REMOVAL OF UNDERARMS, BIKINI, AND FULL LEG WITH THE LUMENIS LIGHTSHEER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGHTSHEER ( ALL MODELS) | OTHER | GEX | LUMENIS, INC. | LIGHTSHEER | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |