LIGHTSHEER ( ALL MODELS)

MDR report key: 2973803 · Received August 7, 2007

Report

Report Number
2914019-2007-00024
Event Type
Other
Date Received
August 7, 2007
Date of Event
July 2, 2007
Report Date
August 7, 2007
Manufacturer
LUMENIS, INC.
Product Code
GEX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER SERVICE TECH, REGULATORY INSPECTION NOTED DIRTY TOUCHSCREEN AND ENERGY METER GLASS. HANDPIECE TIP DIRTY AND LOOKS CLOUDED. PERFORMED INITIAL FINAL AND CONFIRM ENERGY OUTPUT. FIRED AT 37J WHICH WAS AT 40MS. THE DISPLAY GASKET WAS DISLOCATED AND DIRTY, REPLACED THE GASKET. THE ENERGY METER GLASS HAD THICK COATING OF SUBSTANCE. CLEANED TOUCHSCREEN, ENERGY METER GLASS, AND CLEAN TIP. PERFORMED ENERGY METER CALIBRATION. PERFORMED NECESSARY RELIABILITY IMPROVEMENTS AND FINAL TEST. ROOT CAUSE TOUCHSCREEN AND METER GLASS DIRTY.

Description of Event or Problem · 1

ONE PATIENT HAS BEEN BURNED, NAME IS (B)(6) AND I AM THE AESTHETIC NURSE. THE PROCEDURE WAS LASER HAIR REMOVAL OF UNDERARMS, BIKINI, AND FULL LEG WITH THE LUMENIS LIGHTSHEER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGHTSHEER ( ALL MODELS) OTHER GEX LUMENIS, INC. LIGHTSHEER NA

Patients

Seq Age Sex Outcome Treatment
1 Other