TRAPEZOID¿ RX
Report
- Report Number
- 3005099803-2013-00896
- Event Type
- Malfunction
- Date Received
- February 22, 2013
- Date of Event
- January 28, 2013
- Report Date
- January 28, 2013
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- LQC
- PMA / PMN Number
- K040447
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
PATIENT AGE/DATE OF BIRTH IS UNKNOWN; HOWEVER, THE PATIENT WAS REPORTED AS BEING OVER 18 YEARS OLD. (B)(4). THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND RESIDUE ON THE DEVICE INDICATING PROCEDURAL USE AND THE BASKET WAS IN THE CLOSED POSITION WITH THE TIP INTACT. THE GUIDEWIRE LUMEN (SIDE-CAR) PRESENTED PUSH-BACK AND THE COIL ASSEMBLY WAS FOUND TO BE BUCKLED NEAR THE DISTAL END. THE GUIDEWIRE LUMEN WAS TORN AT BOTH THE PROXIMAL AND DISTAL ENDS. A FUNCTIONAL EVALUATION TO EXTEND THE BASKET WAS UNSUCCESSFUL DUE THE RESIDUE INSIDE THE DEVICE. THE BASKET WAS MANUALLY EXTENDED AND RETRACTED AND THE WIRES WERE FOUND TO BE EVEN SPACED AND NOT DEFORMED. THE BASKET WIRES WERE FOUND TO BE COVERED WITH HEAVY GREEN RESIDUE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS CONSISTENT WITH THE COMPLAINT THAT THE BASKET TIP FAILED TO DETACH. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS THE COMPLAINT IS ASSOCIATED WITH A PRODUCT THAT MEETS THE DESIGN AND MANUFACTURE SPECIFICATION BUT DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, PERFORMANCE WAS LIMITED. (B)(4).
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET WAS USED DURING A BILE DUCT STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, AFTER USING THE BASKET DEVICE TO REMOVE SEVERAL STONES WITHOUT ISSUE FROM THE BILE DUCT, THE PHYSICIAN GRASPED AND CAPTURED THE FINAL STONE WITHIN THE BASKET. HOWEVER, THE 2CM STONE APPARENTLY BECAME JAMMED IN THE BASKET, AND AN ATTEMPT TO DISENGAGE THE BASKET'S DISTAL TIP WAS UNSUCCESSFUL. REPORTEDLY, AN ALLIANCE HANDLE WAS USED DURING THE PROCEDURE AND DID NOT MALFUNCTION. THE BASKET WITH THE ENCAPSULATED STONE WAS WITHDRAWN ALONG WITH THE ENDOSCOPE FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET DEVICE. NO DEVICE DAMAGE WAS NOTED.. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED AS BEING STABLE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET WAS USED DURING A BILE DUCT STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, AFTER USING THE BASKET DEVICE TO REMOVE SEVERAL STONES WITHOUT ISSUE FROM THE BILE DUCT, THE PHYSICIAN GRASPED AND CAPTURED THE FINAL STONE WITHIN THE BASKET. HOWEVER, THE 2CM STONE APPARENTLY BECAME JAMMED IN THE BASKET, AND AN ATTEMPT TO DISENGAGE THE BASKET'S DISTAL TIP WAS UNSUCCESSFUL. REPORTEDLY, AN ALLIANCE HANDLE WAS USED DURING THE PROCEDURE AND DID NOT MALFUNCTION. THE BASKET WITH THE ENCAPSULATED STONE WAS WITHDRAWN ALONG WITH THE ENDOSCOPE FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET DEVICE. NO DEVICE DAMAGE WAS NOTED.. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED AS BEING STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 77737 | TRAPEZOID¿ RX | LITHOTRIPTOR, BILIARY MECHANICAL | LQC | BOSTON SCIENTIFIC - SPENCER | M00510890 | 15582486 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |