FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2973702 · Received February 22, 2013

Report

Report Number
1416980-2013-04422
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
January 1, 2013
Report Date
January 31, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED ONE UNIT FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE UNIT SHOWED NO EVIDENCE OF PHYSICAL ABNORMALITY. A LEAK TEST WAS PERFORMED WITH NO EVIDENCE OF A LEAK ANYWHERE ON THE UNIT. CONDENSATION WAS FOUND ON THE UNIT BUT NOT A LEAK. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE INCORRECT LOT NUMBER WAS REPORTED IN THE INITIAL MEDWATCH.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE LEAKED AFTER FILLING. THE DEVICE WAS FILLED WITH A SOLUTION OF PAMIDRONATE. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77691 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 12H041

Patients

Seq Age Sex Outcome Treatment
1 PAMIDRONATE