FDA Adverse Event Malfunction Summary report: N

SMALL BATTERY DRIVE

MDR report key: 2973692 · Received February 22, 2013

Report

Report Number
8030965-2013-00593
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
January 17, 2013
Report Date
January 25, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS DEVICE RECEIVED FOR EVALUATION, DATE UNKNOWN. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED PROBLEM WAS CONFIRMED. THE DEVICE WOULD NOT ENGAGE WHEN POWER WAS APPLIED DURING TESTING. THE UNIT EXHIBITED DAMAGE TO THE MOTOR AND ECU THAT WAS CONSISTENT WITH IMMERSION IN WATER. THIS CONDITION IS INDICATIVE OF IMPROPER CLEANING OF THE DEVICE. WE RECOMMEND THAT THE USER REVIEW THE CLEANING AND MAINTENANCE PROCEDURES FOR THE DEVICE. PROPER CLEANING AND MAINTENANCE ARE REQUIRED IN ORDER TO ENSURE TROUBLE-FREE OPERATION.

Additional Manufacturer Narrative · 1

THE DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN REQUESTED.

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. SHOULD BE EXEMPT: ENTERED IN ERROR.

Description of Event or Problem · 1

THIS REPORT IS 1 OF 1 FOR COMPLAINT (B)(4).

Description of Event or Problem · 1

A REPORT WAS RECEIVED REGARDING A BATTERY DRIVE. IT WAS REPORTED THAT THE DEVICE WORKED INTERMITTENTLY EVEN WHEN NEW BATTERIES WERE INSTALLED. REPORTEDLY, THE DEVICE WAS NOT USED IN A SURGICAL PROCEDURE, THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM TO USER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78487 SMALL BATTERY DRIVE SMALL BATTERY DRIVE HWE SYNTHES GMBH 004366

Patients

Seq Age Sex Outcome Treatment
1