FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 2973672 · Received February 22, 2013

Report

Report Number
3005075853-2013-00794
Event Type
Malfunction
Date Received
February 22, 2013
Date of Event
December 4, 2012
Report Date
February 1, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED WITH THE BLADE BROKEN OFF AND NOT RETURNED. DURING FUNCTIONAL TESTING ON A GENERATOR THE DEVICE GAVE AN ERROR CODE 5. IN ADDITION, THE HAND CONTROL SWITCH ASSEMBLY WAS NOT FUNCTIONAL. HOWEVER, THE DEVICE DID ACTIVATE WHEN TESTED WITH THE FOOT SWITCH. THE DEVICE WAS DISASSEMBLED TO INSPECT INTERNAL COMPONENTS. DRIED BODY FLUIDS WERE FOUND AT THE HAND ACTIVATION DOMES. IT IS PROBABLE THAT THIS MAY HAVE AFFECTED THE FUNCTIONALITY OF THE HAND ACTIVATION BUTTONS.THE DEVICE WILL STOP ACTIVATING, AND DISPLAY AN INSTRUMENT ERROR ON THE GENERATOR WHEN THE BLADE BECOMES DAMAGED. THE LOCATION OF THE BREAK IS INSIDE THE TUBE PROXIMAL TO THE TISSUE PAD. THE ANALYSIS CONCLUDED THAT THE BLADE BROKE OFF DUE TO CONTACT WITH THE INNER TUBE. THIS DAMAGE WAS MOST LIKELY DUE TO EXCESSIVE SIDE FORCES APPLIED TO THE BLADE WHILE ACTIVATING RESULTING IN BLADE CONTACT WITH THE INNER TUBE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TIP BROKE OFF IN THE PATIENT. THE TIP WAS REMOVED AND INCLUDED IN THE RED BAG. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. ONE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77669 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE