FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2973580 · Received February 19, 2013

Report

Report Number
1627487-2013-05229
Event Type
Injury
Date Received
February 19, 2013
Date of Event
January 23, 2013
Report Date
January 23, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CA(B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPG'S AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 4 OF 4. REFERENCE MFR REPORT#: 1627487-2013-05226, 05227, 05228. THE PT HAS TWO IPG'S. IT WAS REPORTED THE PT EXPERIENCES HEATING AT THE IPG SITE WHILE RECHARGING. IT IS UNK WHICH IPG IS ASSOCIATED WITH POCKET HEATING. A REPLACEMENT CHARGING SYSTEM RESOLVED THE PT'S ISSUE. ON (B)(4) 2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PT'S RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PT'S. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72610 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3454860

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention IMPLANT:| IMPLANT:| SCS LEADS: MODEL 3163 (X4)| SCS EXTENSIONS: MODEL 3341 (X2)