FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2973359 · Received February 21, 2013

Report

Report Number
2938836-2013-00183
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
December 6, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

EXTERNAL INSULATION ABRASION WAS FOUND AT 46.0-46.5CM FROM THE TIP ELECTRODE, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT EXTERNALIZED CONDUCTORS WERE OBSERVED BETWEEN THE RV AND SVC COIL. NO ELECTRICAL ANOMALIES WERE NOTED. LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76460 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1 71 YR