FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2973351 · Received February 21, 2013

Report

Report Number
2938836-2013-00181
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
December 12, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 21.2CM WAS RETURNED FOR ANALYSIS. THE PORTION OF THE LEAD THAT WAS RETURNED WAS NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT INAPPROPRIATE SHOCK DUE TO NOISE AND HIGH IMPEDANCE WERE OBSERVED. NO CAPTURE WAS ALSO NOTED. INSULATION ANOMALY, FRACTURE, OR CLAVICULAR CRUSH SUSPECTED. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76070 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/60 NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR