FDA Adverse Event
Malfunction
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 2973351
·
Received February 21, 2013
Report
- Report Number
- 2938836-2013-00181
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Date of Event
- December 12, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 21.2CM WAS RETURNED FOR ANALYSIS. THE PORTION OF THE LEAD THAT WAS RETURNED WAS NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT INAPPROPRIATE SHOCK DUE TO NOISE AND HIGH IMPEDANCE WERE OBSERVED. NO CAPTURE WAS ALSO NOTED. INSULATION ANOMALY, FRACTURE, OR CLAVICULAR CRUSH SUSPECTED. THE LEAD WAS CAPPED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76070 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7000/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |