FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS

MDR report key: 2973336 · Received February 21, 2013

Report

Report Number
2955842-2013-00597
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 22, 2013
Report Date
January 24, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING OBSERVED THAT THE PITCH UP CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL INSTALLED IN CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT WRIST ARE NOT DAMAGED. ADDITIONAL OBSERVATION NOT REPORTED BY SITE IS TUBE DAMAGE. DISTAL END OF MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SCRATCHES ARE .085 - .180 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. EVIDENCE NOT CONCLUSIVE, BUT DAMAGE MAY HAVE BEEN CAUSED BY MISHANDLING/MISUSE. PER THE CUSTOMER, AT THE TIME OF THE REPORTED EVENT, NOTHING FELL INTO A PATIENT DURING A SURGICAL PROCEDURE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVE. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CENTRAL PROCESSING, A BROKEN CABLE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS NOTED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76415 FENESTRATED BIPOLAR FORCEPS ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10121127 496

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES