ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2013-00116
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Date of Event
- January 27, 2013
- Report Date
- January 28, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- MMI
- PMA / PMN Number
- K922823/A007
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
SAMPLES WERE COLLECTED IN 13X100 SERUM SEPARATOR TUBES AND CENTRIFUGED AT 4500 RPMS FOR 5 MINUTES. NO SAMPLE INTEGRITY ISSUES WERE NOTED. ALL SYSTEM PARAMETERS (INCLUDING QC, CALIBRATIONS, AND SYSTEM CHECKS) WERE PERFORMING WITHIN ASSAY/INSTRUMENT SPECIFICATIONS PRIOR TO THE EVENT. ALL LEVELS OF ACCUTNI QC WERE FOUND TO BE ELEVATED AND RECOVERED OUTSIDE OF THE LABORATORY'S ESTABLISHED RANGES AFTER THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO VERIFY HARDWARE PERFORMANCE IN RESPONSE TO THIS EVENT. FSE PERFORMED A SYSTEM CHECK AND ELEVATED RLUS WERE GENERATED. FSE REPLACED THE SUBSTRATE PROBE AND DECONTAMINATED THE SUBSTRATE SYSTEM IN RESPONSE TO THE ELEVATED RELATIVE LIGHT UNITS (RLU) RECOVERY. A SYSTEM CHECK WAS THEN PERFORMED AND WAS FOUND TO BE PASSING WITHIN INSTRUMENT SPECIFICATIONS; THEREFORE, THE FSE DETERMINED THE CAUSE OF THIS EVENT TO BE HARDWARE MALFUNCTION LEADING TO CONTAMINATION OF THE SUBSTRATE SYSTEM. FSE ALSO PROACTIVELY REPLACED INCUBATOR BELT AND CLIPS, THE ASPIRATE PROBES, AND ALL OF THE PERI PUMP TUBING. THE MIXERS, BELT, AND WASH WHEEL WERE INSPECTED AND NO ISSUES WERE NOTED. ALIGNMENTS AND SPEED SETTINGS WERE CHECKED AND NO ADJUSTMENTS WERE MADE. SYSTEM VERIFICATION PROCEDURES WERE FOUND TO BE PASSING WITHIN ASSAY/INSTRUMENT/LABORATORY SPECIFICATIONS FOLLOWING SERVICE. THE LIKELY CAUSE OF THIS EVENT IS ATTRIBUTED TO HARDWARE.
A CUSTOMER CONTACTED BECKMAN COULTER REGARDING ERRONEOUSLY ELEVATED TROPONIN I (ACCUTNI) RESULTS WITHIN THE RISK STRATIFICATION RANGE ON 4 EMERGENCY DEPARTMENT PATIENTS GENERATED BY THEIR ACCESS 2 IMMUNOASSAY SYSTEM. THE ERRONEOUSLY ELEVATED ACCUTNI RESULTS WERE RELEASED OUTSIDE OF THE LABORATORY; HOWEVER, THE CUSTOMER CONFIRMED THERE WAS NO CHANGE TO, OR IMPACT ON, PATIENT TREATMENT WITH REGARD TO THIS EVENT. THE SAMPLES WERE REPEATED ON THE LABORATORY'S ALTERNATE ACCESS 2 ANALYZER AND ACCUTNI RESULTS WITHIN THE NORMAL REFERENCE RANGE FOR EACH PATIENT WERE OBTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76962 | ACCESS® 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | MMI | BECKMAN COULTER | ACCESS 2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |