FDA Adverse Event Injury Summary report: N

SEE H10

MDR report key: 2973166 · Received February 21, 2013

Report

Report Number
2015691-2013-19350
Event Type
Injury
Date Received
February 21, 2013
Date of Event
June 7, 2012
Report Date
February 1, 2013
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MODEL#: THIS DEVICE IS NOT A DISTRIBUTED, MARKETED, OR SOLD IN THE U.S.; HOWEVER, IT IS SIMILAR TO MODEL NO. 2800 - CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS, WHICH IS MARKETED IN THE U.S. DEVICE NOT RETURNED. UNFORTUNATELY, THE VALVE WAS NOT RETURNED TO EDWARDS FOR ANALYSIS, AND THEREFORE, THE ROOT CAUSE OF THIS EVENT COULD NOT BE INVESTIGATED. THE TYPE AND CAUSE OF REGURGITATION VARIES DEPENDING UPON MULTIPLE FACTORS. TYPICALLY, MILD REGURGITATION IS NOT UNUSUAL AFTER INITIAL VALVE REPLACEMENT, AND IS USUALLY TOLERATED BY PATIENTS. OFTEN MODERATE TO HIGH REGURGITATION REQUIRING RE-OPERATION IN THE IMMEDIATE POST-OPERATIVE PERIOD IS DUE TO PROCEDURAL RELATED ISSUES AND IS UNRELATED TO THE DEVICE. REGURGITATION WHICH DEVELOPS PROGRESSIVELY OVER TIME CAN BE DUE TO A NUMBER OF ISSUES INCLUDING PATIENT RELATED FACTORS OR STRUCTURAL VALVE DETERIORATION. HOWEVER, ADVANCES IN VALVE DESIGN AND BIOPROSTHETIC MATERIAL HAVE BEEN MADE WITH THE INTENTION OF REDUCING PERIVALVULAR OR CENTRAL LEAKS BY PROVIDING MORE EFFICIENT HEMODYNAMICS AND LONGER TISSUE DURABILITY. IN THIS CASE, PER THE HEALTH-CARE PROVIDER, THE EVENT WAS PROCEDURE RELATED AND NOT RELATED TO ANY DEVICE MALFUNCTION OR QUALITY DEFICIENCY. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION.

Description of Event or Problem · 1

EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE VALVE WAS EXPLANTED AFTER AND IMPLANT DURATION OF APPROXIMATELY 9 DAYS. BASED ON THE FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, THE EXPLANT WAS DUE TO A PERIVALVULAR LEAK. THE HEALTH-CARE PROVIDER NOTED THAT THE EXPLANT WAS PROCEDURE RELATED AND NOT RELATED TO ANY DEVICE MALFUNCTION. NO OTHER DETAILS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75552 SEE H10 REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES SEE H10

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| R