PRECISION XCEED PRO
Report
- Report Number
- 2954323-2013-00125
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Date of Event
- February 4, 2013
- Report Date
- February 4, 2013
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE REPORTED METER WAS RETURNED AND INVESTIGATED WITH RETAINED TEST STRIPS. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. GIVEN NO TEST STRIPS WERE RETURNED FOR INVESTIGATION, RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (47HT5H) WERE TESTED WITH CONTROL SOLUTION. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED. METHOD: IN ADDITION, THIS SERVES AS A CORRECTION REPORT. HEALTH PROFESSIONAL QUESTION OF THE INITIAL MDR IS ANSWERED INCORRECTLY. HAS BEEN UPDATED TO REFLECT THE CORRECT ENTRY.
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. NOTE: THE DATE OF MANUFACTURE IS UNKNOWN.
A HEALTHCARE PROFESSIONAL REPORTED RECEIVING INACCURATE READINGS ON THEIR ADC BLOOD GLUCOSE METER. THE HEALTHCARE PROFESSIONAL REPORTED COMPARING AND ADC METER READING OF 200 MG/DL TO A LAB RESULT OF 600 MG/DL. THE BLOOD TESTS WERE REPORTEDLY PERFORMED WITHIN TEN MINUTES. THE RESULTS WERE PLOTTED ON A PARKES ERROR GRID AND FELL IN THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES WAS CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76831 | PRECISION XCEED PRO | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 47HT5H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |