FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2973051 · Received February 21, 2013

Report

Report Number
1416980-2013-04410
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 1, 2013
Report Date
January 28, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH AN F-82 ALARM WAS CONFIRMED BY BAXTER SERVICE PERSONNEL. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE A DEFECTIVE CENTRAL PROCESSING UNIT (CPU) BOARD. THE CPU WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP WITH AN F-82 ALARM. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.