FDA Adverse Event Malfunction Summary report: N

BIPOL LEAD MODEL 300

MDR report key: 2972782 · Received February 21, 2013

Report

Report Number
1644487-2013-00476
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
February 4, 2013
Report Date
February 4, 2013
Manufacturer
CYBERONICS INC
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT FLUID LEAKS WERE FOUND IN THE LEAD DURING PRODUCT ANALYSIS OF THE LEAD. THE LEAD WAS RETURNED DUE TO AN INFECTION REPORTED ON MEDWATCH #1644487-2013-00055. THE ABRADED OPENINGS FOUND ON THE OUTER AND INNER SILICONE TUBES AND THE PUNCTURE MARK FOUND ON THE OUTER SILICONE TUBING, MOST LIKELY PROVIDED THE LEAKAGE PATH FOR WHAT APPEARED TO BE REMNANTS OF DRIED BODY FLUIDS FOUND INSIDE THE OUTER AND INNER SILICONE TUBES OF THE LEAD. WHAT APPEARED TO BE WHITE DEPOSITS WERE OBSERVED IN VARIOUS LOCATIONS. EDS (ENERGY DISPERSION SPECTROSCOPY - PROVIDES CHEMICAL OR ELEMENT IDENTITY/COMPOSITION ANALYSIS) WAS PERFORMED AND IDENTIFIED THE DEPOSIT AS CONTAINING PHOSPHORUS, SODIUM, SULPHUR AND CALCIUM. EXCEPT FOR THE ABRADED OPENINGS, THE CONDITION OF THE RETURNED LEAD PORTION IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. WITH THE EXCEPTION OF THE ABRADED TUBING OPENINGS, NO OBVIOUS ANOMALIES WERE NOTED. THE SETSCREW MARKS FOUND ON THE CONNECTOR PINS PROVIDE EVIDENCE THAT, AT ONE POINT IN TIME, A GOOD MECHANICAL AND ELECTRICAL CONNECTION WAS PRESENT. CONTINUITY CHECKS OF THE RETURNED LEAD PORTION WERE PERFORMED, DURING THE VISUAL ANALYSIS, WITH NO DISCONTINUITIES IDENTIFIED. BASED ON THE FINDINGS IN THE PRODUCT ANALYSIS LAB, THERE IS EVIDENCE OF AN ABRADED INNER TUBING OPENING IN THE RETURNED PORTION OF THE DEVICE

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76534 BIPOL LEAD MODEL 300 LEAD LYJ CYBERONICS INC 300-20 16992C

Patients

Seq Age Sex Outcome Treatment
1 24 YR