BIPOL LEAD MODEL 300
Report
- Report Number
- 1644487-2013-00476
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Date of Event
- February 4, 2013
- Report Date
- February 4, 2013
- Manufacturer
- CYBERONICS INC
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
IT FLUID LEAKS WERE FOUND IN THE LEAD DURING PRODUCT ANALYSIS OF THE LEAD. THE LEAD WAS RETURNED DUE TO AN INFECTION REPORTED ON MEDWATCH #1644487-2013-00055. THE ABRADED OPENINGS FOUND ON THE OUTER AND INNER SILICONE TUBES AND THE PUNCTURE MARK FOUND ON THE OUTER SILICONE TUBING, MOST LIKELY PROVIDED THE LEAKAGE PATH FOR WHAT APPEARED TO BE REMNANTS OF DRIED BODY FLUIDS FOUND INSIDE THE OUTER AND INNER SILICONE TUBES OF THE LEAD. WHAT APPEARED TO BE WHITE DEPOSITS WERE OBSERVED IN VARIOUS LOCATIONS. EDS (ENERGY DISPERSION SPECTROSCOPY - PROVIDES CHEMICAL OR ELEMENT IDENTITY/COMPOSITION ANALYSIS) WAS PERFORMED AND IDENTIFIED THE DEPOSIT AS CONTAINING PHOSPHORUS, SODIUM, SULPHUR AND CALCIUM. EXCEPT FOR THE ABRADED OPENINGS, THE CONDITION OF THE RETURNED LEAD PORTION IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. WITH THE EXCEPTION OF THE ABRADED TUBING OPENINGS, NO OBVIOUS ANOMALIES WERE NOTED. THE SETSCREW MARKS FOUND ON THE CONNECTOR PINS PROVIDE EVIDENCE THAT, AT ONE POINT IN TIME, A GOOD MECHANICAL AND ELECTRICAL CONNECTION WAS PRESENT. CONTINUITY CHECKS OF THE RETURNED LEAD PORTION WERE PERFORMED, DURING THE VISUAL ANALYSIS, WITH NO DISCONTINUITIES IDENTIFIED. BASED ON THE FINDINGS IN THE PRODUCT ANALYSIS LAB, THERE IS EVIDENCE OF AN ABRADED INNER TUBING OPENING IN THE RETURNED PORTION OF THE DEVICE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76534 | BIPOL LEAD MODEL 300 | LEAD | LYJ | CYBERONICS INC | 300-20 | 16992C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |