ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-02008
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Report Date
- January 30, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 (B)(4) -DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP'S HISTORY SHOWED NO RECORD OF PUMP NOT PRIMED OR NO CARTRIDGE DETECTED WARNINGS. DURING TESTING, THE PUMP SUCCESSFULLY PASSED THE REWIND, LOAD, AND PRIME STEPS. THE FORCE SENSOR WAS FOUND TO BE IN CALIBRATION. THE PUMP WAS OPENED AND NO DAMAGE WAS FOUND TO THE FORCE SENSOR CIRCUIT BOARD.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE. REPORTEDLY, THE PUMP WAS DISPENSING INSULIN DURING THE LOAD STEP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 77104 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |