FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 2972740
·
Received February 21, 2013
Report
- Report Number
- 1416980-2013-04367
- Event Type
- Malfunction
- Date Received
- February 21, 2013
- Date of Event
- November 12, 2012
- Report Date
- November 12, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE REPORTED PROBLEM OF A FLOGARD INFUSION PUMP THAT IS "NOT WORKING" WAS CONFIRMED IN SAMPLE EVALUATION TASK. FAILURE F49 WAS FOUND DURING EVALUATION WHICH WAS DUE TO CPU SETTING CONFIGURATION WERE OUT OF SPECIFICATION. SERVICE RESET CPU PARAMETERS TO RESOLVE THE REPORTED PROBLEM.
Description of Event or Problem · 1
THE FACILITY REPORTED A FLOGARD INFUSION PUMP THAT "IS NOT WORKING". IT IS UNKNOWN IF THIS EVENT OCCURRED DURING DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO REPORT OF PATIENT/USER INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 77412 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |