FDA Adverse Event Malfunction Summary report: N

ASSY, PROBE, BLADDERSCAN BVI 9400/9600

MDR report key: 2972612 · Received February 15, 2013

Report

Report Number
3022472-2013-00019
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 18, 2013
Report Date
January 18, 2013
Manufacturer
VERATHON MEDICAL
Product Code
IYO
PMA / PMN Number
K071217
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE SERVICE REPRESENTATIVE REPLACED THE FAULTY DCM. HE ALSO FOUND THE PROBE'S PCBA J7 PIN WAS DAMAGED AND REPLACED THE PROBE PCBA. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE DISPLAYED LOW READINGS. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69428 ASSY, PROBE, BLADDERSCAN BVI 9400/9600 NONE IYO VERATHON MEDICAL 0570-0188

Patients

Seq Age Sex Outcome Treatment
1