FDA Adverse Event Malfunction Summary report: N

STERRAD NX STERILIZER

MDR report key: 2972492 · Received February 21, 2013

Report

Report Number
2084725-2013-00074
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 29, 2013
Report Date
January 29, 2013
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K042116
Removal / Correction Number
Z-0744-0746-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. THE VACUUM PUMP, CATALYTIC DECOMP FILTER, AND OIL MIST FILTER WERE REPLACED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE ON (B)(4) 2013.

Additional Manufacturer Narrative · 1

CONCLUSION: BASED ON CAPA INVESTIGATION AND ASTM TESTING, OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD SYSTEMS. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS (FMEA), AND SYSTEM HAZARD AND USER MISUSE ANALYSIS (SHUMA), AND CORRECTIVE AND PREVENTATIVE ACTION (CAPA). THE DHR WAS REVIEWED. THE UNIT MET MANUFACTURING SPECIFICATIONS AT THE TIME OF RELEASE AND THERE WERE NO ISSUES RELATED TO THIS FAILURE MODE NOTED. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST SIX MONTHS (JULY 2012 THROUGH JANUARY 2013) DID NOT OBSERVE ANY SIGNIFICANT TREND. THE TREND FOR THE PRODUCT MALFUNCTION CODE OF ODOR/SMELLS WAS COMPLETED FROM JANUARY 2013 THROUGH DECEMBER 2013 AND REVEALED A SIGNIFICANT TREND WHICH IS ADDRESSED IN CAPA. THE FMEA REVEALED THE RISK PRIORITY NUMBER (RPN) SCORES ARE BELOW 100 AND IS CONSIDERED ACCEPTABLE. THE SHUMA INDICATES THE RISK IS BROADLY ACCEPTABLE FOR MODERATE LEVEL INJURY AND AS LOW AS REASONABLY PRACTICABLE FOR MILD (LIMITED) EXPOSURE TO ODOR OR ODORANTS. CAPA INVESTIGATION AS WELL AS THE ASTM TESTING PERFORMED INDICATED OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD® SYSTEMS. THE CATALYTIC CONVERTER WAS RETURNED AND TESTED. THE CATALYTIC CONVERTER PASSED A PUMP DOWN TEST AND EXHIBITED NO ODOR. THE REASON FOR RETURN OF THE CATALYTIC CONVERTER WAS NOT CONFIRMED. THE OIL MIST FILTER WAS RETURNED AND TESTED. THE OIL MIST FILTER PASSED A PUMP DOWN TEST AND HAD NO ODOR DETECTED. THE REASON FOR RETURN OF THE OIL MIST FILTER WAS NOT CONFIRMED. THE VACUUM PUMP WAS RETURNED AND TESTED. THE VACUUM PUMP WAS UNABLE TO PASS A PUMP DOWN TEST AND THERE WAS AN ODOR DETECTED. THE REASON FOR RETURN OF THE VACUUM PUMP WAS CONFIRMED. THE REPORTED ISSUE WAS RESOLVED AT THE CUSTOMER FACILITY. THE ISSUE WILL BE TRACKED AND TRENDED. NO FURTHER INVESTIGATION IS NECESSARY AT THIS TIME.

Description of Event or Problem · 1

A CUSTOMER REPORTED AN EVENT OF A "STRONG ODOR" EMITTING FROM THE STERRAD NX STERILIZER. THERE WAS NO REPORT OF HUMAN REACTION. CUSTOMER WAS ADVISED TO TURN THE UNIT OFF AND VACATE THE AREA. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75956 STERRAD NX STERILIZER STERRAD EQUIPMENT (MLR) MLR ADVANCED STERILIZATION PRODUCTS NA