FDA Adverse Event Malfunction Summary report: N

ISOLA SET SCW HEXDRIVER

MDR report key: 2972397 · Received February 21, 2013

Report

Report Number
1526439-2013-12689
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 29, 2013
Report Date
January 29, 2013
Manufacturer
DEPUY SYNTHES SPINE
Product Code
LXH
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL EXAMINATION CONFIRMED THE HEX TIP OF THE DRIVER HAD BEEN SNAPPED OFF. REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE PERFORMED AS THE LOT CODE IS UNKNOWN. REVIEW OF PRODUCT COMPLAINTS FOUND NO EMERGING TRENDS. ALTHOUGH NO DEFINITIVE CONCLUSIONS CAN BE MADE, THE OBSERVED DAMAGE SUGGESTS THAT THE DEVICE HEX TIP MOST LIKELY BROKE OFF DUE TO AN UNANTICIPATED TORSIONAL STRESS. IN THE ABSENCE OF AN OBSERVED TREND, THIS COMPLAINT WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED.

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED UPON RECEIPT OF THE DEVICE AND COMPLETION OF THE EVALUATION.

Description of Event or Problem · 1

INTERNATIONAL AFFILIATE REPORTS THAT RETURNED LOAN KIT INSPECTION FOUND THE TIP OF THE SET SCREW HEX DRIVER HAD BROKEN OFF FROM THE INSTRUMENT. IT IS NOT KNOWN WHEN OR WHERE THE TIP BREAKAGE OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76485 ISOLA SET SCW HEXDRIVER ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH DEPUY SYNTHES SPINE

Patients

Seq Age Sex Outcome Treatment