PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE
Report
- Report Number
- 1419937-2013-00103
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 17, 2013
- Report Date
- January 17, 2013
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. IN F/U WITH THE CUSTOMER. IN F/U WITH THE CUSTOMER, THE CUSTOMER INDICATED THAT THE HOUSING FOR THEIR TRANSFORMER CAME APART, EXPOSING THE UNDERLYING ELECTRONICS. THE CUSTOMER ALSO STATED THAT SHE WOULD RETURN THE ORIGINAL PRODUCT. AS OF THE DATE OF THIS REPORT THE ORIGINAL PRODUCT HAS NOT BEEN RECEIVED. SHOULD THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED OR CORRECTED INFO, A F/U REPORT WILL BE FILED AT THAT TIME. THOUGH THE PRODUCT HAS NOT BEEN RECEIVED AND EVALUATED, THIS TYPE OF FAILURE IS TYPICALLY ASSOCIATED WITH DROPPING THE UNIT UNTO A HARD SURFACE OR OTHER TYPE OF SEVERE IMPACT. THE ORIGINAL DESIGN TESTING INVOLVED DROPPING THE UNIT FROM A 1.0 M HEIGHT PER UL2601-1 2ND EDITION. PRODUCTION SAMPLES WERE DROPPED THREE TO FIVE TIMES FROM A HEIGHT OF 1.0 M AND THE HOUSINGS SHOWED DAMAGE AND CRACKING (ONLY AFTER AT LEAST THREE DROPS). THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING INVESTIGATED (B)(4) IN ORDER TO DETERMINE ROOT CAUSE AND WILL CONTINUE TO BE MONITORED.
THE CUSTOMER REPORTED THAT THE HOUSING ON THEIR PUMP IN STYLE TRANSFORMER WAS FALLING APART, EXPOSING THE UNDERLYING ELECTRONICS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 69496 | PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE | HGX | MEDELA, INC. | 57063 / 9207010 | 2011 / REV L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |