FDA Adverse Event Malfunction Summary report: N

ULTRAFLEX¿ TRACHEOBRONCHIAL

MDR report key: 2972078 · Received February 21, 2013

Report

Report Number
3005099803-2013-01117
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 29, 2013
Report Date
January 29, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
JCT
PMA / PMN Number
K012883
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) FOR THE REPORTED ISSUE OF STENT COVER MISSING. THE COMPLAINANT INDICATED THAT THE DEVICE REMAINS IMPLANTED AND WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE COMPLAINANT PROVIDED AN IMAGE OF THE DEVICE POST DEPLOYMENT. A REVIEW OF THE IMAGE BY BSC COULD NOT DEFINITIVELY CONCLUDE IF THE COVER WAS MISSING, AS THE COVER IS TRANSLUCENT AND DIFFICULT TO VISUALIZE ONCE WET. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED, A SUPPLEMENTAL MEDWATCH WILL BE FILED

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX TRACHEOBRONCHIAL COVERED STENT WAS IMPLANTED WITHIN THE RIGHT BRONCHUS OF A PATIENT ON (B)(6) 2013 DURING A STENT PLACEMENT PROCEDURE. ACCORDING TO THE COMPLAINANT, THE INDICATION FOR THE STENT PLACEMENT WAS FOR TREATMENT OF A STENOSIS CAUSED BY A TUMOR WITHIN THE RIGHT BRONCHUS. THE PATIENT'S ANATOMY WAS REPORTED TO NOT BE TORTUOUS AND HAD NOT BEEN DILATED PRIOR TO STENT PLACEMENT. NO VISIBLE ISSUES WERE NOTED TO THE DEVICE PRIOR TO THE PROCEDURE. DURING THE PROCEDURE, AN ULTRAFLEX TRACHEOBRONCHIAL COVERED STENT WAS SUCCESSFULLY IMPLANTED WITHIN THE RIGHT BRONCHUS. HOWEVER, FOLLOWING DEPLOYMENT, IT WAS NOTICED THAT THE STENT COVER WAS MISSING. NO INTERVENTION WAS PERFORMED AS A RESULT OF THIS EVENT. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED TO BE STABLE POST PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77173 ULTRAFLEX¿ TRACHEOBRONCHIAL PROSTHESIS, TRACHEAL, EXPANDABLE JCT BOSTON SCIENTIFIC - GALWAY M00564750 15648535

Patients

Seq Age Sex Outcome Treatment
1 60 YR