FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2972059 · Received February 21, 2013

Report

Report Number
2032227-2013-00698
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
February 8, 2013
Report Date
February 8, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BLOOD GLUCOSE LEVELS GREATER THAN 600 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE CUSTOMER INSULIN PUMP PASSED THE PRIME TEST. THE CUSTOMER DIDN'T HAVE A TUBING CLAMP FOR THE HIGH PRESSURE TEST, BUT DID RECEIVE A NO DELIVERY ALARM EARLIER DUE TO A BENT CANNULA. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS FLUSH. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75628 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAB

Patients

Seq Age Sex Outcome Treatment
1 73 YR