FDA Adverse Event Malfunction Summary report: N

MODULAR HEXLOBE DRIVER X15

MDR report key: 2971996 · Received February 21, 2013

Report

Report Number
1526439-2013-12661
Event Type
Malfunction
Date Received
February 21, 2013
Report Date
February 5, 2013
Manufacturer
DEPUY SYNTHES SPINE
Product Code
LXH
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED DRIVER CONFIRMED DISTAL TIP BREAKAGE. AN OPTICAL IMAGE OF THE FRACTURED SURFACE REVEALED PLASTIC DEFORMATION AND RATCHET MARKINGS STRETCHING IN A CIRCULAR PATTERN AROUND THE CENTER OF THE IMAGE. THIS SUGGESTS THE DRIVER UNDERWENT A STATIC FAILURE DUE TO SHEAR TORSIONAL OVERLOADING WHILE TIGHTENING. SCANNING ELECTRON MICROSCOPY CONFIRMED THAT BREAKAGE LIKELY OCCURRED DUE TO HIGH TORSIONAL SHEAR LOADS. NO MATERIAL DEFECTS OR OTHER ABNORMALITIES WERE OBSERVED IN THIS ANALYSIS. REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DISCREPANCIES. AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.

Description of Event or Problem · 1

INTERNATIONAL AFFILIATE REPORTS INSPECTION OF A RETURNED LOAN KIT FOUND THE TIP OF THE HEXLOBE DRIVER HAD BROKEN OFF FROM THE INSTRUMENT. IT IS NOT KNOWN WHEN/WHERE TIP BREAKAGE OCCURRED. THIS REPORT IS BEING REPORTED LATE. THE NEED TO REPORT MDRS WITHIN THIRTY DAYS OF BECOMING AWARE OF THE EVENT HAS BEEN REVIEWED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76592 MODULAR HEXLOBE DRIVER X15 ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH DEPUY SYNTHES SPINE G0808

Patients

Seq Age Sex Outcome Treatment
1