Description of Event or Problem · 1

MEDICAL ASSISTANT WORKING IN BRONCHOSCOPY PROCEDURE ROOM OBSERVED WHAT APPEARED TO BE FOREIGN SUBSTANCE STAINING INSIDE OF STAINLESS STEEL BIOPSY PORT RESERVOIR OF BRONCHOSCOPE. DARK RED RESIDUE CAME OFF ON COTTON-TIPPED APPLICATOR. UPON EXAMINING ENTIRE INVENTORY OF BRONCHOSCOPES IN MPU (MEDICAL PROCEDURE UNIT), SIMILAR RESIDUE/STAINING WAS DISCOVERED IN FIVE (5) OTHER BRONCHOSCOPES. THE INITIAL BRONCHOSCOPE WAS REWASHED AND HIGH-LEVEL DISINFECTED; AND HAD NO FURTHER VISIBLE RESIDUE/STAINING AFTER SAME. IT WAS PUT BACK INTO CIRCULATION. THE REMAINING FIVE SCOPES WERE SEQUESTERED FOR MANAGEMENT FOLLOW-UP. AFTER ADDITIONAL REPROCESSING, ONLY ONE OF THE FIVE BRONCHOSCOPES HAD NO FURTHER VISIBLE RESIDUE/STAINING. REPORTER VALIDATED THAT THE CORRECT BRUSH IS BEING USED IN POST-PROCEDURE MANUAL CLEANING PROCESS; AND THAT STAFF VERBALIZED APPROPRIATE STEPS OF MANUAL CLEANING PROCESS.WHAT WAS THE ORIGINAL INTENDED PROCEDURE?BRONCHOSCOPIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57965 OLYMPUS BRONCHOSCOPE (FLEXIBLE OR RIGID) KOG OLYMPUS AMERICA, INC BF-UC180 UNK
57966 OLYMPUS BRONCHOSCOPE (FLEXIBLE OR RIGID) KOG OLYMPUS AMERICA, INC. BF-UC180 UNK
57967 OLYMPUS BRONCHOSCOPE (FLEXIBLE OR RIGID) KOG OLYMPUS AMERICA, INC. BF1T180 UNK
57968 OLYMPUS BRONCHOSCOPE (FLEXIBLE OR RIGID) KOG OLYMPUS AMERICA, INC. BF1T180 UNK
57969 OLYMPUS BRONCHOSCOPE (FLEXIBLE OR RIGID) KOG OLYMPUS AMERICA, INC. BF3C160 UNK