Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE CINE BRIDGE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A "CINE DISK NOT AVAILABLE" ERROR UPON BOOT UP RESULTING IN A LOSS OF SUBTRACTION, ROAD-MAPPING, OR OTHER CRITICAL VASCULAR CINE FUNCTIONS. THERE IS NO REPORT OF PT INJURY.