FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2971756 · Received February 6, 2013

Report

Report Number
9680959-2013-00307
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
January 24, 2013
Report Date
February 6, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND CORRECTLY SET THE BIOS SETTINGS. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE FIELD ENGINEER REPORTED THAT THE SYSTEM WAS NOT SAVING PT IMAGES. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
51669 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1