FDA Adverse Event Malfunction Summary report: N

COMPLETE MULTIPURPOSE EASY RUB

MDR report key: 2971731 · Received January 30, 2013

Report

Report Number
3004537773-2013-00006
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
ADVANCED MEDICAL OPTICS INC
Product Code
LPN
PMA / PMN Number
K014202
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MANUFACTURING RECORDS FOR THE REPORTED LOT WAS PERFORMED; NO DEVIATIONS WERE NOTED AND PRODUCT MET ALL SPECIFICATIONS. FUNCTIONAL EVAL OF THE PRODUCT RETAINED FROM LOT AK01123 WERE PERFORMED; ALL RESULTS WERE WITHIN SPECIFICATION. VISUAL EXAM OF THE PRODUCT RETAINED FROM THE REPORTED LOT WERE PERFORMED; ALL RESULTS WERE WITHIN SPECIFICATION. VISUAL EVAL OF THE RETURNED DEVICE WAS PERFORMED. THE FLIP-TOP CAP SHOWED A PIN HOLE IN THE FRONT SIDE OF THE BOTTLE CAP. FURTHER INSPECTION OF THE HOLE SHOWED THAT THE BOTTLE HAD BEEN SCRATCHED. THE DIRECTION OF THESE SCRATCHES FOES FROM OUTSIDE TO INSIDE. IT APPEARS THAT THE CAP WAS PIERCED BY A NEEDLE. THIS WAS CONFIRMED WHEN WE FOUND SCRATCHES INSIDE THE FLIP TP CAP. IT APPEARS THAT THE UNIT WAS MANIPULATED AFTER IT LEFT THE MANUFACTURING SITE. ALL PERTINENT INFORMATION HAS BEEN PROVIDED.

Description of Event or Problem · 1

CONSUMER REPORTED THAT HER BOTTLE OF COMPLETE MULTIPURPOSE SOLUTION EASY RUB HAD A SMALL HOLE IN THE CAP AND LEAKED. REPORTER DID NOT USE THE PRODUCT AND RETURNED IT TO THE MFR FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41735 COMPLETE MULTIPURPOSE EASY RUB LPN ADVANCED MEDICAL OPTICS INC AK01123

Patients

Seq Age Sex Outcome Treatment
1