FDA Adverse Event Other Summary report: N

BMC RADIOFREQUENCY PUNCTURE GENERATOR

MDR report key: 2971588 · Received June 29, 2007

Report

Report Number
9710452-2007-00011
Event Type
Other
Date Received
June 29, 2007
Report Date
June 29, 2007
Manufacturer
BAYLIS MEDICAL CO., INC.
Product Code
GEI
PMA / PMN Number
K013904
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PROBLEM OCCURRED BECAUSE OF USER ERROR. THE RADIOFREQUENCY GENERATOR BEGINS TO DELIVER RF ENERGY WHEN THE SWITCH IS TURNED ON. THE RF ENERGY CAN BE SET TO BE DELIVERED FOR THE APPROPRIATE DESIRED TIME. THE DELIVERY OF THE RF ENERGY CAN BE DISCONTINUED BEFORE THE SET TIME IS REACHED BY SWITCHING OFF THE RADIOFREQUENCY GENERATOR. THESE DIRECTIONS ARE ADEQUATELY DESCRIBED IN THE USER MANUAL OF THE RADIOFREQUENCY GENERATOR. THE PT WOLD HAVE UNDERGONE OPEN HEART SURGERY IF THE INTERVENTION WITH THE RADIOFREQUENCY WIRE AND BMC RADIOFREQUENCY PUNCTURE GENERATOR WOULD NOT HAVE BEEN ATTEMPTED OR IF THE INTERVENTION WAS ATTEMPTED UNSUCCESSFULLY. SINCE THIS PROBLEM OCCURRED BY USER ERROR, NO IMMEDIATE CORRECTIVE ACTION WILL BE TAKEN.

Description of Event or Problem · 1

A PT WAS BEING TREATED FOR PULMONARY ATRESIA USING THE NYKANEN RADIOFREQUENCY WIRE IN CONJUNCTION WITH THE BMC RADIOFREQUENCY PUNCTURE GENERATOR. ONCE THE NYKANEN WIRE CROSSED THE SEPTUM, THE PHYSICIAN INSTRUCTED THE NURSE TO DISCONTINUE THE RF ENERGY FROM THE GENERATOR. THE NURSE ASSUMED THAT THE RF ENERGY WOULD BE DISCONTINUED WHEN THE SWITCH IS RELEASED. SHE WAS NOT AWARE THAT SHE NEEDED TO SWITCH THE GENERATOR OFF. THE WIRE CONTINUED TO ADVANCE WHILE RF ENERGY WAS STILL BEING DELIVERED, ULTIMATELY RESULTING IN PERFORATION OF THE MAIN PULMONARY ARTERY. THE PT HAD TO BE TAKEN FOR SURGERY TO CORRECT THIS PERFORATION. THE SURGERY WAS DONE WITHOUT COMPLICATIONS. THE PT WILL BE FOLLOWED UP AS PER THE USUAL POST-OPERATIVE CARE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BMC RADIOFREQUENCY PUNCTURE GENERATOR ELECTROSURGICAL GENERATOR GEI BAYLIS MEDICAL CO., INC. REP-100-115 012004-025

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention