FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2971507 · Received February 4, 2013

Report

Report Number
1720753-2013-01414
Event Type
Malfunction
Date Received
February 4, 2013
Date of Event
January 24, 2013
Report Date
February 4, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE LEMO CONNECTOR AND INTERCONNECT CABLE WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY DISPLAYED AN X-RAY DISABLED ERROR MESSAGE. THIS ERROR WOULD PREVENT FLUOROSCOPY X-RAY FROM BEING PERFORMED. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47361 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1