FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2971485
·
Received February 4, 2013
Report
- Report Number
- 1720753-2013-01402
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- January 22, 2013
- Report Date
- February 4, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. A FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A HIGH MA ERROR MESSAGE, BUT THAT THEY WERE UNABLE TO CONTINUE WITH THE CASE. THERE IS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47274 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |