FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2971470 · Received February 4, 2013

Report

Report Number
8020893-2013-00277
Event Type
Malfunction
Date Received
February 4, 2013
Date of Event
October 10, 2012
Report Date
January 10, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK DISPLAY WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47354 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1