FDA Adverse Event Malfunction Summary report: N

SCREWDRIVER HEX 2.5MM AO FITTING

MDR report key: 2971405 · Received February 21, 2013

Report

Report Number
0008031020-2013-00019
Event Type
Malfunction
Date Received
February 21, 2013
Date of Event
January 28, 2013
Report Date
January 28, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-SELZACH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE AFFECTED DEVICE WAS RETURNED FOR INVESTIGATION. THE DEVICE LOOKS WORN DOWN FROM USE. THE AFFECTED DEVICE WAS SENT TO THE MANUFACTURING CELL FOR DIMENSIONAL AND FUNCTIONAL INSPECTION. RESULTS SHOW THE DEVICE IS OUT OF SPECIFICATION DUE TO THE FACT IT WAS USED MULTIPLE TIMES. THE AO COUPLING IS ALSO DAMAGED. THEREFORE, THE REPORTED EVENT THAT THE SCREWDRIVER SLIPS IN THE SCREW HEAD COULD BE CONFIRMED. BASED ON INVESTIGATION, THE ROOT CAUSE OF THE DETERMINED DAMAGE WAS ATTRIBUTED TO A USE RELATED ISSUE (WEAR AND TEAR). THE DEVICE WAS MANUFACTURED MORE THAN 3 YEARS AGO. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. INDICATIONS FOR ANY MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE NOT DETERMINED IN THE INVESTIGATION.

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

2.5 HEX SCREWDRIVERS IN MATTA PELVIC BASIC INSTRUMENTS ARE WORN DOWN FROM USE AND CAUSED SCREWDRIVERS TO SLIP IN SCREW HEAD WHILE TURNING.

Description of Event or Problem · 1

2.5 HEX SCREWDRIVERS IN MATTA PELVIC BASIC INSTRUMENTS ARE WORN DOWN FROM USE AND CAUSED SCREWDRIVERS TO SLIP IN SCREW HEAD WHILE TURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77037 SCREWDRIVER HEX 2.5MM AO FITTING INSTRUMENT LXH STRYKER OSTEOSYNTHESIS-SELZACH NA

Patients

Seq Age Sex Outcome Treatment
1 Other