GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2013-00161
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- November 20, 2012
- Report Date
- November 21, 2012
- Manufacturer
- AMERICAN MEDICAL SYSTEMS
- Product Code
- GEX
- PMA / PMN Number
- K100746
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
FIBER ANALYSIS: THE FIBER'S METAL AND GLASS CAPS WERE FOUND TO BE DETACHED; THE FIBER BROKEN PROXIMAL TO THE FIBER/CAP FUSION ZONE; THE METAL AND GLASS CAPS WERE NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER/CAP CONDITIONS WOULD RESULT IN FORWARD FIRING. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION/ USER HANDLING DUE TO TISSUE CONTACT AND OR ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, DAMAGE TO THE FIBER CAP WAS OBSERVED AT 182,918 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47532 | GREENLIGHT MOXY FIBER OPTIC | POWERED SURGICAL LASER INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS | 10-2400 | 229A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| GREENLIGHT XPS LASER SYSTEM |