FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 2971376 · Received February 4, 2013

Report

Report Number
2937094-2013-00161
Event Type
Malfunction
Date Received
February 4, 2013
Date of Event
November 20, 2012
Report Date
November 21, 2012
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
PMA / PMN Number
K100746
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FIBER ANALYSIS: THE FIBER'S METAL AND GLASS CAPS WERE FOUND TO BE DETACHED; THE FIBER BROKEN PROXIMAL TO THE FIBER/CAP FUSION ZONE; THE METAL AND GLASS CAPS WERE NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER/CAP CONDITIONS WOULD RESULT IN FORWARD FIRING. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION/ USER HANDLING DUE TO TISSUE CONTACT AND OR ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, DAMAGE TO THE FIBER CAP WAS OBSERVED AT 182,918 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47532 GREENLIGHT MOXY FIBER OPTIC POWERED SURGICAL LASER INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS 10-2400 229A

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| GREENLIGHT XPS LASER SYSTEM