FDA Adverse Event Injury Summary report: N

PUMP MMT-722NAP PRDGM INS V2.2 PL EN

MDR report key: 2971316 · Received April 23, 2008

Report

Report Number
3004209178-2008-09734
Event Type
Injury
Date Received
April 23, 2008
Date of Event
April 8, 2008
Report Date
April 11, 2008
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TESTS. THE DISPLAY WAS MISSING SEGMENTS DUE TO A FAULTY LCD DISPLAY.

Description of Event or Problem · 1

THE CUSTOMER STATED, SHE WAS HOSPITALIZED DUE TO HYPERGLYCEMIA. THE BLOOD GLUCOSE READING AT THE TIME OF THE EVENT WAS NOT REPORTED. IT WAS REPORTED THAT THE DISPLAY WAS MISSING SEGMENTS. TROUBLESHOOTING WAS NOT PERFORMED FOR THE HYPERGLYCEMIA BECAUSE OF THE DISPLAY ANOMALY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAP PRDGM INS V2.2 PL EN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 31 YR Hospitalization