FDA Adverse Event Injury Summary report: N

GENTLEPOWR LUX CONTRA ANGLE 25LPA

MDR report key: 2971237 · Received June 20, 2007

Report

Report Number
1419798-2007-00018
Event Type
Injury
Date Received
June 20, 2007
Date of Event
April 30, 2007
Report Date
May 15, 2007
Manufacturer
KAVO AMERICA
Product Code
EFB
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

MAINTENANCE INFO WAS REVIEWED WITH THE OFFICE AND MAINTENANCE SHEETS WERE SENT. HEAD, COVER NUT AND BACK END CAP ARE DAMAGED DUE TO DENTS AT 10 AND 12, BEARINGS/GEARS ARE WORN AND GRITTY IN DRIVE ASSEMBLY AND SHAFT DUE TO THE DENTS. DENTS IN HEAD CAUSES ATTACHMENT TO HEAT UP, THERE WAS MEDIUM AMOUNT OF DEBRIS FOUND IN UNIT.

Description of Event or Problem · 1

THE HANDPIECE BECAME HOT AND BURNED THE PT'S TONGUE. ONE PART MAALOX, ONE PART BENADRYL ELIXIR AND ONE PART XYLOCAINE VISCUS WAS PRESCRIBED AS NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWR LUX CONTRA ANGLE 25LPA OPERATIVE DENTAL UNIT ACCESSORY EFB KAVO AMERICA 25LPA

Patients

Seq Age Sex Outcome Treatment
1 UNK