FDA Adverse Event
Injury
Summary report: N
GENTLEPOWR LUX CONTRA ANGLE 25LPA
MDR report key: 2971237
·
Received June 20, 2007
Report
- Report Number
- 1419798-2007-00018
- Event Type
- Injury
- Date Received
- June 20, 2007
- Date of Event
- April 30, 2007
- Report Date
- May 15, 2007
- Manufacturer
- KAVO AMERICA
- Product Code
- EFB
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
MAINTENANCE INFO WAS REVIEWED WITH THE OFFICE AND MAINTENANCE SHEETS WERE SENT. HEAD, COVER NUT AND BACK END CAP ARE DAMAGED DUE TO DENTS AT 10 AND 12, BEARINGS/GEARS ARE WORN AND GRITTY IN DRIVE ASSEMBLY AND SHAFT DUE TO THE DENTS. DENTS IN HEAD CAUSES ATTACHMENT TO HEAT UP, THERE WAS MEDIUM AMOUNT OF DEBRIS FOUND IN UNIT.
Description of Event or Problem · 1
THE HANDPIECE BECAME HOT AND BURNED THE PT'S TONGUE. ONE PART MAALOX, ONE PART BENADRYL ELIXIR AND ONE PART XYLOCAINE VISCUS WAS PRESCRIBED AS NEEDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWR LUX CONTRA ANGLE 25LPA | OPERATIVE DENTAL UNIT ACCESSORY | EFB | KAVO AMERICA | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |