FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2971200 · Received February 1, 2013

Report

Report Number
1720753-2013-01376
Event Type
Malfunction
Date Received
February 1, 2013
Date of Event
January 23, 2013
Report Date
February 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE DISPLAY BOARD POWER SUPPLY VOLTAGE WAS ADJUSTED. THE SYS WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS NO IMAGE ON THE MONITOR AND THE SYS WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
45343 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1