FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2971177 · Received February 1, 2013

Report

Report Number
9680959-2013-00287
Event Type
Malfunction
Date Received
February 1, 2013
Date of Event
January 24, 2013
Report Date
February 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CONNECTORS WERE RESEATED, THE POWER SUPPLY WAS REPLACED, AND THE SOFTWARE UPGRADE WAS INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY SYSTEM WOULD NOT BOOT UP. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
45516 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900