FDA Adverse Event
Malfunction
Summary report: N
PLUM 1.6 W/ DATAPORT
MDR report key: 2971132
·
Received February 1, 2013
Report
- Report Number
- 9615050-2013-00173
- Event Type
- Malfunction
- Date Received
- February 1, 2013
- Date of Event
- January 1, 2013
- Report Date
- January 14, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K911401
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RECEIVED ON (B)(4) 2013. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
DURING VERIFICATION TESTING AT THE SERVICE CENTER, CHARRING WAS NOTED ON THE STRAIN OF FEMALE END OF THE AC POWER CORDS. THE OUTER INSULATION OF THE AC POWER CORD HAD SEPARATED FROM THE FEMALE END OF THE AC POWER CORD AND THE INTERNAL INSULATION OF THE WIRES WERE BROKEN WITH EXPOSED WIRES. AT THE MALE END OF THE AC POWER CORD, THE GROUND PRONG WAS FOUND TO BE BROKEN OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 45204 | PLUM 1.6 W/ DATAPORT | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |