FDA Adverse Event Malfunction Summary report: N

PLUM 1.6 W/ DATAPORT

MDR report key: 2971132 · Received February 1, 2013

Report

Report Number
9615050-2013-00173
Event Type
Malfunction
Date Received
February 1, 2013
Date of Event
January 1, 2013
Report Date
January 14, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K911401
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED ON (B)(4) 2013. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING VERIFICATION TESTING AT THE SERVICE CENTER, CHARRING WAS NOTED ON THE STRAIN OF FEMALE END OF THE AC POWER CORDS. THE OUTER INSULATION OF THE AC POWER CORD HAD SEPARATED FROM THE FEMALE END OF THE AC POWER CORD AND THE INTERNAL INSULATION OF THE WIRES WERE BROKEN WITH EXPOSED WIRES. AT THE MALE END OF THE AC POWER CORD, THE GROUND PRONG WAS FOUND TO BE BROKEN OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
45204 PLUM 1.6 W/ DATAPORT 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA